Customize Configuration
12-50 Vivid S70 / S60 – User Manual
BC092760-1EN
01
DICOM environment or if any exam at any later time is
expected to be recalled to Vivid S70 / S60 or EchoPAC
Software Only using DICOM Query/Retrieve. If the
DICOM server does not allow the private data format
within DICOM SR files, this feature should be disabled.
• No images: no images are sent, only M&A.
• Signed Doppler velocities: send signed Doppler
velocities.
• Use older SR version: when checked a Use older SR
version pull-down menu is displayed. The current exam
data will be sent in the same format as the selected SR
version. Details about format and content of the SR
version can be found in the corresponding user manual
of the selected version.
These settings apply to both “TID 5200 Echocardiography
Procedure Report” and “TID 5100 Vascular Ultrasound
Procedure Report”
6. Select OK.
DICOM spooler
DICOM spooler displays the current DICOM output jobs. The
jobs may be Storage, Print, Modality Performed Procedure Step
or Storage Commitment. The DICOM spooler is used for
checking the current job's status when a job is saved or when
the total spooler status on the right of the Archive window
displays an error.
From the DICOM spooler the user can also:
• Delete non-active jobs.
• Resend a job that has failed or is in hold.
• Send a job that has failed or is in hold, to a new destination.
• Hold a job that is not active.