Connectivity configuration
Vivid S70 / S60 – User Manual 12-51
BC092760-1EN 01
The job's status displayed in the DICOM spooler window can be:
• Pending: the job is complete, waiting to be active.
• Hold: the job is complete, but suspended, waiting for an
action from the user.
• Append: the job is incomplete, waiting for more images
(Direct store function).
• Active: the job is complete and connected to the destination
device.
• Failed: the job is complete but one or more images failed to
transmit to the destination device.
• Done: the job is saved to the destination device. The jobs
that are done are removed from the spooler after a few
minutes.
To start the DICOM spooler:
1. Do one of the following:
• Press Utility and Spooler on the control panel
• Press F4 on the alphanumeric keyboard
• Press Alt + S on the alphanumeric keyboard
The DICOM spooler window is displayed (see
Figure 12-35).
The DICOM spooler window is automatically updated. Press
Refresh to update the information displayed at any time.
Figure 12-35. The DICOM job spooler window
To delete a job:
1. Select the job(s) to delete in the DICOM job spooler window.
NOTE: Only non-active jobs can be deleted.
2. Press Delete.
To resend a job:
1. Select the job(s) to re-send in the DICOM job spooler
window.