User Manual corpuls
3
Maintenance and tests
ENG - Version 2.1 – P/N 04130.2 211
Functional
test
Description Measures by the user Correct result
Temperature
measurement
Function
check of
temperature
measurement
Connect a temperature
sensor to the patient box.
If the temperature value is
not displayed in a
parameter field, it may be
not configured.
Select the parameter field
in which the value should
be shown
Take the temperature
sensor into your hand.
Display of the room
temperature.
The temperature measuring
value increases when the
temperature sensor is in your
hand.
Non-invasive
blood pressure
monitoring
Function
check of non-
invasive blood
pressure
monitoring
Perform a blood pressure
measurement on a
volunteer subject with the
NIBP cuff.
If the non-invasive blood
pressure value is not
displayed in a parameter
field, it may be not
configured.
Select the parameter field
in which the value should
be shown
The blood pressure is displayed
in a parameter field.
Invasive blood
pressure
monitoring
Ability to
calibrate the
transducer
Test the ability to calibrate
the transducer. To do this,
expose the transducer to
atmospheric pressure.
In the main menu, select
under "IBP" ► "Calib.
P {measuring value}".
The invasive blood pressure is
displayed
Function
check of
invasive blood
pressure
measurement
Various transducers have
equipment for a functional
test. Perform the functional
test according to the
operating instructions for
the transducer.
If the measured value of
invasive blood pressure is
not displayed in a
parameter or curve field, it
may be not configured.
Select the parameter or
curve field in which the
value should be shown
An invasive blood pressure is
displayed in the parameter or
curve field after the functional
test is complete.
The displayed pressure curve
shows scaling.