Procedure in case of malfunctions
User Manual corpuls
3
236 ENG - Version 2.1 P/N 04130.2
Alarm message Priority Explanation/Measure
Connect therapy
electrodes
High priority
Connect the corPatch electrodes or shock paddles
to the therapy master cable of the corpuls
3
.
If the alarm message persists, connect immediately
the reserve corPatch electrodes.
CPR cable loose
CPR cable loose (P-
Box)
Medium priority The cable indicated is not connected to the patient
box.
Check the respective cable and reconnect if
necessary.
CPR sensor activ.
CPR sen. activ.(P-Box)
Medium priority The sensor has been activated.
CPR sensor expired
CPR sen. exp. (P-Box)
Medium priority The sensor is expired and has to be exchanged with
a new one.
CPR sensor loose
CPR sensor loose (P-
Box)
Medium priority The sensor indicated has come loose from the
intermediate cable.
Check the sensor and reconnect if necessary.
CPR sensor not
activated
CPR n. activ.(P-Box)
Medium priority The sensor has not been activated.
For correct use, the sensor has to be activated.
CPR sensor-NOT FOR
PAT. USE
CPR-NO PAT. USE (P-
Box)
Medium priority Using this sensor on a patient is prohibited.
Contact authorised sales and service partners.
D-ECG transfer to
server failed
Low priority Connection to server or to patient box is interrupted.
Restart connection.
Repeat recording and transmission of D-ECG.
Defective Oxi adhesive
sensor
Def. Oxi adhes. sensor
(P-Box)
Medium priority The adhesive sensor is defective or has expired.
The sensor LED's have to be aligned directly over
each other at the measuring site.
Defective Oxi cable
Medium priority The intermediate cable is defect or not connected
correctly.
Disconnect intermediate cable and connect again.
If the sensor LED does not flash, the intermediate
cable has to be exchanged.
Defective Oxi sensor
Defect. Oxi sen. (P-Box)
Medium priority The sensor is defect or not connected correctly.
Disconnect sensor and connect again.
If the sensor LED does not flash, the sensor has to
be exchanged.
Defib temperature high
(X)
High priority „X“ stands for a failure number from 1 to 2.
The defibrillator was very frequently charged within a
short period.
Avoid too frequent internal discharges.
Defibrillator alarm High priority
The corpuls
3
still functions correctly, but should
however be checked by a service technician as soon
as possible.
Contact authorised sales and service partners.