Procedure in case of malfunctions
ENG - Version 2.1 P/N 04130.2 237
Alarm message Priority Explanation/Measure
Defibrillator failure (3)
High priority Internal error.
The corpuls
3
is possibly not functioning correctly
and must not be used.
Contact authorised sales and service partners.
Defibrillator failure (x) High priority X“ stands for a failure number from 1 to 7.
The corpuls
3
is possibly not functioning correctly
and must not be used.
Contact authorised sales and service partners.
ECG cable (4 pole)
loose
4-pole cable loose (P-
Box)
Medium priority Check if the 4-pole ECG cable is connected correctly
to the patient box.
ECG electrode (X)
loose
Electr. (X) loose (P-box)
Medium priority „X“ stands for a an electrode number from V1 to V6.
Connect the clip of the complementary 6 pole ECG
diagnostic cable to the loose ECG electrode or
disconnect the ECG cables not used from corpuls
3
(see chapter 6.3.2 Preparing the Patient for a D-
ECG, page 103).
Check ECG electrodes.
Make sure that all clips are correctly connected to
the contacts and/or the ECG electrodes are perfectly
placed on the patient’s skin.
ECG electrode R/RA
loose
Electr. R/RA loose (P-
box)
ECG electrode L/LA
loose
Electr. L/LA loose (P-
box)
ECG electrode F/LL
loose
Electr. F/LL loose (P-
box)
Medium priority Connect the clip of the ECG monitoring cable to the
loose ECG electrode (see chapter 6.3.2 Preparing
the Patient for a D-ECG, page 103). Check ECG
electrodes.
Make sure that all clips are correctly connected to
the contacts and/or the ECG electrodes are perfectly
placed on the patient’s skin.
ECG electrodes loose
ECG electr. loose (P-
Box)
Medium priority More than one ECG electrode are not connected to
the patient.
Check ECG electrodes.
ECG failure (X)
ECG failure (X) (P-Box)
High priority „X“ stands for a failure number from 1 to 5.
The corpuls
3
is possibly not functioning correctly
and must not be used.
Contact authorised sales and service partners.
Error CPR module
Error CPR (P-Box)
Medium priority The measuring option mentioned is faulty.
Contact authorised sales and service partners.
Error CPR sensor
Error CPR sen. (P-Box)
Medium priority The sensor mentioned is faulty and has to be
replaced.
Contact sales and service partners, if necessary.