Procedure in case of malfunctions
User Manual corpuls
3
238 ENG - Version 2.1 – P/N 04130.2
Alarm message Priority Explanation/Measure
Low priority The GSM module is faulty.
Contact authorised sales and service partners.
Error IBP module Medium priority The measuring option mentioned is faulty.
Contact authorised sales and service partners.
Low priority The insurance card was not recognised or could not
be read.
Take the insurance card out of the corpuls
3
and
check if it was inserted with the chip oriented
towards the front.
Only German, Austrian and Swiss insurance cards
and Belgian ID-cards are supported.
The reading process may take up to 10 seconds.
NIBP HW error (P-box)
Medium priority Wait up to 20 secs. to start measurement. The
softkeys are greyed out as long as the selection is
not available.
If the patient is not in a critical condition, restart the
corpuls
3
.
Check if the patient as a palpable pulse.
Error T1/T2 Medium priority The measuring channel mentioned is defect.
Contact authorised sales and service partners.
Low priority The fax machine of the recipient is busy.
Try to repeat fax transmission at a later time.
Fax connection not
possible
Low priority Possibly a wrong fax number has been dialled.
Repeat dialing.
Low priority Fax transmission was not successful.
Repeat procedure
Contact authorised sales and service partners, if
necessary.
GPRS authorization
failed
Low priority No GPRS service available for this SIM card.
Wrong APN configured or connection with the GSM
network aborted.
Repeat attempt to establish connection.
Low priority The APN is not configured or has not been saved.
Configure APN anew and save configuration.
Contact your mobile communications provider for the
respective current APN.
Low priority The reception quality is too low. No connection to
the mobile communications network can be
established.
If possible, select a location with better reception.
If necessary, repeat procedure at a later time.
HR > [NUMBER]/min
High priority The measured heart rate exceeds/falls below the
upper/lower alarm limit.
Check the patient's vital signs.