Procedure in case of malfunctions
User Manual corpuls
3
240 ENG - Version 2.1 – P/N 04130.2
Alarm message Priority Explanation/Measure
Low perf. (X) (P-Box)
Low priority "X" stands for the vital parameters SpCO, SpHb or
SpMet.
The measured signal is too weak.
Make sure that the patient lies calmly during the
measurement and that there is no commotion
caused by the vehicle.
Select another measuring site.
High priority Alarm message shown on the display of the patient
box or the monitoring unit.
The respective module restarts due to a software
error.
(not available)
High priority Check if the module is switched on.
Wireless communication between the defibrillator
and monitoring unit/patient box is cut off or could not
be established.
Make sure that the distance between the modules
does not exceed 10 m and that no obstacles impair
the connection. Use corpuls
3
as a compact device
if necessary.
Communication in the connected state between the
defibrillator and monitoring unit/patient box is cut off
or could not be established
Check if one of the two infrared interfaces is covered
or heavily soiled.
failed
NIBP automeas. failed
(P-Box)
Medium priority Perform manual measurement.
If the error persists, contact sales and service
partner.
NIBP cuff overpr. (P-
Box)
Medium priority Make sure that the patient does not lie on the hose
or on the cuff.
Make sure that the correct cuff size is used.
Make sure that the hose is not bent or pinched.
Make sure that the cuff is attached correctly.
mmHg
NIBP DIA > [NUMBER]
mmHg
Medium priority The measured diastolic blood pressure exceeds or
falls below the upper/lower alarm limit.
Check the patient's vital signs.
impossible
NIBP no inflation (P-
Box)
Medium priority Make sure that the hose is not bent or pinched.
Make sure that the patient does not lie on the hose
or on the cuff.
Make sure that the cuff is attached correctly.