Procedure in case of malfunctions
ENG - Version 2.1 – P/N 04130.2 241
Alarm message Priority Explanation/Measure
aborted
NIBP measurem.
aborted(P-Box)
Medium priority Wait up to 20 secs. to start measurement. The
softkeys are greyed out as long as the selection is
not available.
If the patient is not in a critical condition, restart the
corpuls
3
.
Check if the patient as a palpable pulse.
NIBP measurement
failed
NIBP meas failed(P-
box)
Medium priority Check if
the cuff hose is properly connected,
the cuff hose is not bent,
artifacts have occurred during measurement,
the NIBP cuff was properly applied or another
application error is present
Repeat measurement.
Medium priority Make sure that the patient lies calmly during the
measurement and that there is no commotion
caused by the vehicle.
Make sure that the cuff is attached correctly.
Make sure that the correct cuff size is used.
signal
NIBP noisy sig. (P-Box)
Medium priority Make sure that the patient lies calmly during the
measurement and that there is no commotion
caused by the vehicle.
Make sure that the cuff is attached correctly.
Make sure that the correct cuff size is used.
Apply the cuff directly at the arm.
NIBP n. a. (P-Box)
Medium priority If the patient is not in a critical condition, restart the
corpuls
3
.
Check if the patient as a palpable pulse.
Perform manual measurement.
NIBP not calibr. (P-Box)
Medium priority The measuring option mentioned is not calibrated.
Contact authorised sales and service partners.
blockage
NIBP pneum block (P-
Box)
Medium priority Make sure that the patient does not lie on the hose
or on the cuff.
Make sure that the correct cuff size is used.
Make sure that the hose is not bent or pinched.
Make sure that the cuff is attached correctly.
NIBP safety off (P-Box)
Medium priority Check the status of the patient.
Try measurement again.
Make sure that the patient lies calmly during the
measurement and that there is no commotion
caused by the vehicle.
Make sure that the cuff is attached correctly.
mmHg
NIBP SYS > [NUMBER]
mmHg
Medium priority The measured systolic blood pressure exceeds or
falls below the upper/lower alarm limit.
Check the patient's vital signs.