Procedure in case of malfunctions
User Manual corpuls
3
242 ENG - Version 2.1 – P/N 04130.2
Alarm message Priority Explanation/Measure
Low priority The APN is not configured or has not been saved.
Configure APN anew and save configuration.
Contact your mobile communications provider for the
respective current APN.
No batt. inserted (Defib)
No battery inserted
(Monitor)
No battery inserted (P-
Box)
High priority The battery in the respective module is missing.
Insert a battery and/or connect the modules together
to be able to call on the energy reserves of the other
modules if the corpuls
3
is to be operated as a
mobile device.
No connection to Defib
Defibrillator N/A
Check if the module is switched on.
Wireless communication between the defibrillator
and monitoring unit/patient box is cut off or could not
be established.
Make sure that the distance between the modules
does not exceed 10 m and that no obstacles impair
the connection.
Use corpuls
3
as a compact device if necessary.
Communication in the connected state between the
defibrillator and monitoring unit/patient box is cut off
or could not be established
Check if one of the two infrared interfaces is covered
or heavily soiled.
If necessary, use corpuls
3
with wireless
connection.
Contact sales and service partners, if necessary.
Check if the module is switched on.
Wireless communication between the defibrillator
and monitoring unit/patient box is cut off or could not
be established.
Make sure that the distance between the modules
does not exceed 10 m and that no obstacles impair
the connection. Use corpuls
3
as a compact device
if necessary.
Communication in the connected state between the
defibrillator and monitoring unit/patient box is cut off
or could not be established
Check if one of the two infrared interfaces is covered
or heavily soiled.
High priority
The corpuls
3
is in AED mode with the shock
paddles connected and the 4 pole ECG monitoring
cable is not connected to the patient box.
Connect the 4-pole ECG monitoring cable.