User Manual corpuls
3
Appendix
ENG - Version 2.1 – P/N 04130.2 295
Description/Explanation
Measuring range 70 to 150 compressions per minute
1.9 cm to 10.16 cm [0.75 inches – 4.0 inches]
Measurement interval Continuous
Operating temperature (sensor) -10°C to +60°C
Storage temperature (sensor) -30°C to +65°C
Relative humidity (sensor) Up to 93% (without condensation)
Storage humidity (sensor) Up to 93% (without condensation)
Sensor interface Type BF, defibrillation proof
Sensor dimensions (H x W) 87 mm x 51 mm [3.4 x 2.0 inches]
Sensor weight (with cable) 50.2 g
Sensor weight (without cable) 28.8 g
Accuracy:
± 3 compressions per minute
± 0.635 cm [0.25 inches]
A-23 CPR Feedback
GSM/GPRS Modem (Optional)
Description/Explanation
Function principle Tri-Band GSM/GPRS 900/1800/1900 MHz
-24 GSM/GPRS Modem (Optional)
Oximetry (Option SpO
2
, SpCO, SpHb, SpMet, manufacturer, Masimo)
Description/explanation
Amplifier Type BF, insulated > 5 kV, defibrillation proof
Alarm Lower limit 65 %
Updating frequency of display
(Oximetry and pulse frequency)
1 Hz
Band width 0.5 Hz to 6 Hz
Measuring range SpO
2
: 1% to 100%
PP: 25/min to 240/min
Calibration range 70% to 99%
Calibration Calibration by reference measurements by means of fractional
saturation measuring on pulse oximetric haemoglobin oxygen
saturation with dyshaemoglobin-free blood
Definition SpO
2
: 1 %
PP: 1/min
SpCO: 1 %
SpMet: 0.1 %
SpHb: 0.1 g/dl or 0.1 mmol/l
Accuracy of the oxygen saturation
± 2% (70% to 100%)
± 3% (50% to 69%)