When setting the amplitude, MINimum, 
MAXimum and DEFault can be used. The range 
depends on the function being used and the output 
termination (50Ω or high impedance). The default 
amplitude for all functions is 100 mVpp (50Ω). 
 
If the amplitude has been set and the output 
termination is changed from 50Ω to high 
impedance, the amplitude will double. Changing 
the output termination from high impedance to 
50Ω will half the amplitude. 
 
Vrms, dBm or Vpp units can be used to specify the 
output unit to use with the current command. The 
VOLT:UNIT command can be used to set the units 
when no unit is specified with the Apply 
command. If the output termination is set to high 
impedance, dBm units cannot be used. The units 
will default to Vpp.  
 
The output amplitude can be affected by the 
function and unit chosen. Vpp and Vrms or dBm 
values may have different maximum values due to 
differences such as crest factor. For example, a 
5Vrms square wave must be adjusted to 3.536 
Vrms for a sine wave. 
The offset parameter can be set to MINimum, 
MAXimum or DEFault. The default offset is 0 
volts. The offset is limited by the output amplitude 
as shown below. 
|Voffset| < Vmax – Vpp/2 
If the output specified is out of range, the 
maximum offset will be set.