104
Open Correction
5
Touch the EXEC key.
Touch when you wish to cancel correction.
(The display will return to the screen shown
in step 2, and the open correction value will
remain unchanged.)
If correction has never been performed, the
correction values become 0.
Correction starts.
Touch when you wish to cancel correction.
(The display will return to the screen shown
in step 2, and the open correction value will
remain unchanged.)
The time required to acquire correction
values varies with the measurement fre-
quency and the number of points.
The next screen will be displayed once cor-
rection has completed normally.
Correction results
(Conductance, susceptance)
Correction
No.
Measurement frequency
• You can check the conductance and sus-
ceptance for each correction point with
the
keys.
• Correction can be performed for imped-
ances of at least 1 k
Ω
. If the impedance
in the open state is less than 1 k
Ω
, an
error will result.
6
Touch the EXIT key twice.
Displays the measurement screen.
• If correction does not complete normally:
(p. 109)
• To disable the correction value: (p. 111)
For spot correction, the correction is valid when the measurement frequency and the spot correc-
tion frequency match.