EasyManua.ls Logo

Hioki IM3536 - X84; Initializing (System Reset)

Hioki IM3536
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
228
Troubleshooting
When no apparent cause can be established
Perform a system reset.
Most settings will revert to their default values at the time the instrument was shipped from the factory.
(Some settings will not revert to their default values. See "Appx. 11 Initial Settings Table" (p. Appx.15).)
A system reset can also be performed with the
*
RST
and
:PRESet
communications commands.
For more information, see the descriptions of the
*
RST
and
:PRESet
communications commands
on the bundled LCR Application Disc.
Initializing (System Reset)
Screen display method (For more information, see p. 24):
(Measurement screen) SET key>(SET screen) ADVANCED tab>RESET key
1
Disconnect the measurement sample.
2
Touch the RESET key.
Touch to cancel the system reset. (The mea-
surement screen will be displayed.)
Settings will revert to their factory defaults,
and the measurement screen will be dis-
played.
If you are unable to display the initialization screen or perform a system reset, perform a full reset.
(See p. 229)

Table of Contents

Related product manuals