Setting Measurement Conditions (basic settings)
User-congurable settings
Measurement speed (AC/DC)
The measurement speed can be set. The slower the measurement speed is, the more accurate are
the results.
Screen display method (For more information, see p. 24.):
AC measurement (Measurement screen) SET key>(SET screen) BASIC tab>SPEED key
DC measurement (Measurement screen) SET key>(SET screen) Rdc tab>SPEED key
1
Select the measurement speed.
Measurement
speed
Measurement
time
Measurement
accuracy
FAST
Short Low
MED
SLOW
SLOW2
Long High
2
Touch the EXIT key twice.
Displays the measurement screen.
• Measurement time varies with the measurement conditions. (See "10.7 About Measurement Times and
Measurement Speed" (p. 217).)
• The waveform averaging function allows you to set the measurement speed at a higher level of detail.
• The measurement speed cannot be set using the SPEED key when the waveform averaging function is
enabled.
(See "Waveform averaging function (increasing measurement precision or measurement speed)" (p. 82).)