235
Discarding the Instrument
11.4 Discarding the Instrument
The instrument uses a lithium-ion battery as a power supply for saving the measurement condi-
tions. When disposing of this instrument, remove the lithium battery and dispose of battery and
instrument in accordance with local regulations.
WARNING
• To avoid electric shock, turn off the power switch and disconnect the power
cord and measurement cable, probe, or xture before removing the lithium bat-
tery.
• Battery may explode if mistreated. Do not short-circuit, recharge, disassemble
or dispose of in re.
• Keep batteries away from children to prevent accidental swallowing.
Lithium Battery Removal
Required tools: One Philips screwdriver (No.1), One tweezers (to remove the lithium battery)
1
Verify that the power is off, and remove
the connection cables and power cord.
2
Remove the eight screws from the sides
and one screw from the rear.
Lithium bat-
tery
Battery
holder
3
Remove the cover.
4
Insert the tweezers between the battery
and battery holder as shown in the dia-
gram below and lift up the battery.
CAUTION
Take care not to short the + and -.
Doing so may cause sparks.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains Perchlorate Material - special
handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Maintenance and Service
11