192
Environmental and Safety Specications
10.2 Environmental and Safety Specications
Operating temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature and
humidity
-10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation)
Operating environment Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Rated supply voltage 100 V AC to 240 V AC (Voltage uctuations of ±10% from the rated supply
voltage are taken into account.)
Rated power supply fre-
quency
50 Hz/ 60 Hz
Maximum rated power 50 VA
Dimensions Approx. 330W×119H×230D mm (12.99" W ×4.69" H ×9.06" D)
(excluding protrusions)
Mass Approx. 4.2 kg (148.1 oz.)
Standards EMC EN61326 Class A
Safety EN61010
Effect of radiated radio-
frequency electromagnetic
eld
Z: ±5% rdg.,
θ
: ±5° at 10 V/m
Effect of conducted radio-
frequency electromagnetic
eld
Z: ±5% rdg.,
θ
: ±5° at 3 V
Dielectric strength Between the power wire and ground wire: 1.62 kV AC for 1 minutes, cutoff
current 10 mA
Product warranty period 1 year

Table of Contents