100
Open Correction
All correction
Simultaneously acquire the open correction values for all measurement frequencies.
Screen display method (For more information, see p. 27):
(Measurement screen) ADJ key>(ADJ screen) OPEN key
1
Touch the ADJUST key.
2
Touch the ALL key and then touch the
EXIT key.
1
2
The OPEN dialog box will be closed, and
the previous correction value will be dis-
played. (If correction has never been per-
formed, the correction values become 0.)
Check that the measurement cable is in an
open circuit state.
3
Touch the EXEC key.
Touch when you wish to cancel
correction. (The display will
return to the screen shown
in Step 2 of the process, and
the open correction value will
remain unchanged.)
The correc-
tion range can
be limited.
(See p.
101)
Correction will start.
Correction value acquisition time:
Approx. 50 seconds
Touch when you wish to cancel correction.
(The display will return to the screen shown
in step 2, and the open correction value will
remain unchanged.)
The next screen will be displayed once cor-
rection has completed normally.
Correction results
(Conductance, susceptance)
Correction
No.
Measurement frequencies
• You can scroll the screen with the
keys.
• Correction can be performed for imped-
ances of at least 1 k
Ω
. If the impedance
in the open state is less than 1 k
Ω
, an
error will result.
4
Touch the EXIT key twice.
Displays the measurement screen.
• If correction does not complete normally:
(p. 109)
• To disable the correction value: (p. 111)