74
Judging Measurement Results
Conguring BIN function settings (judging measured values based on
multiple judgment standards)
Set the upper and lower limit values for two parameters and display up to 10 classications of judg-
ment results. You can also output the judgment results to an external device.
When the measurement result
falls within the BIN category
range, the green indicator will
light up.
When the measurement
result is OUT OF BINS, the
red indicator will light up.
(Judgment result indication LEDs)
BIN2 (numeral) When BIN judgment
- - -
When BIN is not set
OUT OF BINS When does not match any BIN
(Measurement screen)
About BIN measurement
Bin measurement is performed using a process such as the example shown below.
Judgment BIN
number
No. 1
Parameter
No. 3
Parameter
BIN Judgment
results
Judgment Judgment
Judgment Judgment
Judgment Judgment
Judgment Judgment
The instrument will display the rst BIN number for which the measurement value is judged to be
within the set judgment standard.
If none of the BIN judgments is determined to apply, OUT OF BINS will be displayed.
The BIN number for which the measured
value was rst judged to be within the set
judgment standards will be displayed as the
judgment result.
Judgment reference