75
Judging Measurement Results
The BIN decision mode can be set as one of the following:
Absolute value (ABS) setting (p. 76)
Upper limit value HI
IN
Lower limit value
LO
Set absolute values for the upper limit and lower limit val-
ues of the measurement parameters.
The measurement values displayed are the same as
those of the measurement parameters.
Percent (%) Setting (p. 77)
Upper limit value [%] HI
Reference value
IN
IN
Lower limit value [%]
LO
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit
*1
values.
The measurement values displayed are the same as
those of the measurement parameters.
Deviation Percent (
Δ
%)
*2
Setting (p. 77)
Upper limit value [
%] HI
Reference value
IN
IN
Lower limit value [
%]
LO
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit
*1
values.
The measurement values are displayed in deviations (
Δ
%)
from the reference value.
*1: The following equation is used to calculate the comparison upper limit value and comparison lower limit
value.
(In the case of the comparison lower limit value, if a value that is lower than the reference value is set, the
minus (-) sign is required for the percentage setting value.)
= reference value + |reference value| ×
100
Percentage set value
Upper limit comparison value
(Lower limit comparison value)
*2: The following equation is used to calculate the
% value.
Δ
%=
|reference value|
measurement value - reference value
×100
For more information about HI/IN/LO judgment procedures, see p. 70.
By starting with a restrictive standard and setting a series of judgment standards that grow progressively
more permissive, as shown in the gure to the right, you can rank measured elements.
For a BIN number that does not require a BIN judgment, set the upper and lower limit values to OFF.
The measurement conditions that are used when normal measurement
*
is performed are inherited as is for
the measurement conditions when BIN is performed.
BIN measurement can be used even if only the upper or lower limit value has been set.
When only an upper limit
value has been set
Upper
limit
value
HI
IN
When only a lower limit value
has been set
Lower
limit value
IN
LO
*: Indicates measurement
when not using the com-
parator function or BIN
measurement function.
3
Performing Measurements in LCR Mode