77
Judging Measurement Results
Percentage setting and deviation percentage setting
Set the value after setting the judgment mode (p. 69) to BIN.
1
Touch the BIN key on the measurement
screen.
2
Touch the Z key.
The key display differs depending
on the measurement parameter
3
Touch the % key (persentage setting) or
∆% key (deviation percentage setting)
4
Touch the REF key, and enter the refer-
ence value with the numeric keypad.
2
1
Settable range: -9.99999 G to 9.99999 G
5
Touch the ENTER key to conrm the ref-
erence value.
6
Touch the EXIT key.
7
Use the  key to select the BIN num-
ber to set, and touch EDIT key.
1
2
8
Touch the HI key for the No. 1 param-
eter.
No. 1 parameter
No. 3 parameter
3
Performing Measurements in LCR Mode