Judging Measurement Results
9
Enter the upper limit value with the nu-
meric keypad.
Settable range -999.999% to 999.999%
When you do not want to set the upper and
lower limit values, touch the OFF key.
10
Touch the ENTER key to conrm the up-
per limit value.
The display will return to the screen shown
in step 8.
11
Touch the LO key for the No. 1 parameter,
set the lower limit value with the numeric
keypad, and then touch the ENTER key.
Settable range: -999.999% to 999.999%
The display will return to the screen shown
in step 8.
12
Using the same procedure, set the up-
per and lower limit values for the No. 3
parameter.
13
Touch the EXIT key twice.
Displays the measurement screen.
The set reference value and upper and lower
limit values are common to percentage set-
ting and percentage deviation setting.