43
Setting Measurement Conditions (basic settings)
3.4 Setting Measurement Conditions (basic set-
tings)
(There are two types of measurement: AC measurement and DC measurement (p. 40).
The measurement conditions set for AC measurement and DC measurement differ.
Required: Be sure to set.
Optional: Change setting as necessary.
Setting
During AC
measurement
(When the parameter
is other than Rdc)
During DC
measurement
(When the parameter
is other than Rdc)
Ref. Overview
Measurement
frequency
Required - p. 44
Congure based on the mea-
surement sample.
Measurement range Required Required p. 45
Measurement signal
level
Required - p. 49
Line frequency - Required p. 53
Set to the power supply’s fre-
quency.
Measurement speed Optional Optional p. 54
When you wish to perform mea-
surement more quickly: FAST
When you wish to perform mea-
surement at a higher level of
precision: SLOW, or SLOW2
Low Z high accuracy
mode
Optional
(default value: OFF)
Optional
(default value: OFF)
p. 55
To perform high-precision mea-
surement, set to ON.
To measure at high speed: OFF
Average
Optional
(default value: OFF)
Optional
(default value: OFF)
p. 56
To limit variability of displayed
values, set to ON.
Limit
Optional
(default value: OFF)
- p. 58
To limit the voltage or current
that is applied to the sample, set
to ON.
DC bias
Optional
(default value: OFF)
- p. 59
To superimpose a DC voltage on
the measurement signal during
measurement, set to ON.
DC adjust -
Optional
(default value: OFF)
p. 60
To reduce measurement error:
ON
To measure at high speed: OFF
Trigger Synchronous
Output
Trigger Synchronous
delay
*
Optional
(default value: OFF, default value: 0.0010 s)
p. 64
To apply the signal during mea-
surement only, set to ON.
Optional (default value: 0.0010 s)
Set to a sufciently large value
when you wish to stabilize mea-
surement.
DC delay
*
-
Optional
(default value: 0 s)
p. 61
Adjustment delay
*
-
Optional
(default value: 0.0030 s)
p. 62
Trigger
Optional (Default setting: INT)
Measurement is repeated automatically.
p. 62
To input signals and commands
from an external source, set to
EXT.
Trigger delay
*
Optional (default value: 0 s) p. 63
If the trigger function is enabled,
set to a large enough value that
measurement can stabilize.
*Delay time (For more information about delay timing, see "Measurement and data acquisition timing" (p. 65).