119
Load Correction (Correcting Values to Match Reference Values)
To reset the correction condition settings
This section describes how to clear all settings for the selected correction condition number.
Screen display method (For more information, see p. 27):
(Measurement screen) ADJ key>(ADJ screen) LOAD key
1
Select the correction condition number
to reset with the  keys and then
touch the EDIT key.
2
1
2
Touch the RESET key and then touch
the SET key.
Touch when you wish to cancel the reset.
(This dialog box will be closed.)
1
2
3
Touch the EXIT key twice.
Displays the measurement screen.
When load correction fails to complete normally
If correction fails, a window like the following appears. Touch EXIT to close the window and then
congure the correction conditions again.
5
Error Correction