118
Load Correction (Correcting Values to Match Reference Values)
Setting reference values
Enter the reference value for the parameter displayed to the left of the parameter mode for REF1
and the reference value for the parameter displayed to the right of the parameter mode for REF2.
1
Touch the REF1 key.
2
Enter the reference value with the nu-
meric keypad and touch a unit key to
accept the setting.
2
Uses a multiple of ×1.
(Touching EXIT without
touching a unit key will
also cause a multiple of
×1 to be used.)
Touch when you wish
to cancel input. (This
dialog box will be
closed.)
1
Settable range:
Same as the maximum display range for
the selected parameter.
(See "10.1 General Specications"
(p. 187))
If you make a mistake, touch the C key to
reenter the value.
3
Touch the EXIT key.
The dialog box will be closed.
4
Touch the REF2 key and set the refer-
ence value in the same manner.
When DC is selected for the correction frequency setting, only reference value 1 can be set.