Setting Application Settings
Waveform averaging function (increasing measurement precision or
measurement speed)
The number of measurement waveforms for each frequency band is set for the measurement
speed settings (FAST, MED, SLOW, SLOW2), and this function allows you to set the number of
measurement waveforms for each frequency band. Having more waveforms increases the mea-
surement precision, while having fewer waveforms increases the measurement speed.
Screen display method (For more information, see p. 24):
(Measurement screen) SET key>(SET screen) ADVANCED tab>WAVE NUM key
1
Touch the ON key.
2
Select the frequency band for which you
wish to change the number of measure-
ment waveforms with the  key , and
touch the EDIT key.
1
2
Reverts to the number of measurement
waveforms for each measurement speed.
3
Select the number of measurement
waveforms with  key and touch the
EXIT key.
21
(For more information about the valid set-
ting range, see the table on the following
page.)
4
Touch the EXIT key twice.
Displays the measurement screen.
When the waveform averaging function is enabled, the measurement speed cannot be set with the SPEED
key.