15 Templates
900 Touch Control
■■■■■■■■
165
Min. slope
Minimum slope of the pH electrode.
for very good electrodes:
Input range 0.1 to 999.9 %
Default value 96.5 %
for good electrodes:
Input range 0.1 to 999.9 %
Default value 96.0 %
for usable electrodes:
Input range
0.1 to 999.9 %
Default value 95.0 %
Max. slope
Maximum slope of the pH electrode.
for very good electrodes:
Input range 0.1 to 999.9 %
Default value 101.0 %
for good electrodes:
Input range 0.1 to 999.9 %
Default value 102.0 %
for usable electrodes:
Input range
0.1 to 999.9 %
Default value 103.0 %
Response time
The voltage measured after three minutes in stirred solution serves as a
comparison value for determining the response time. The response time is
the time after which the measured voltage is within ±1 mV of this com-
parison value.
for very good electrodes:
Input range 0 to 999 s
Default value 45 s
for good electrodes:
Input range 0 to 999 s
Default value 50 s