28 Manual control
900 Touch Control
■■■■■■■■
259
Proceed as follows:
1
Configure the dosing function
NOTICE
■ The dosing and filling rates should be decreased for viscous liq-
uids.
■ The maximum dosing and filling rates depend on the cylinder
volume (see chapter 32, page 492).
■ Enter the dosing rate.
■ Enter the filling rate.
2
Start dosing
■ Tap on [Dose].
The volume dosed is displayed. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.
3
Fill the cylinder
■ Tap on [Fill].
The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.
NOTICE
If you exit the dialog with the fixed key [ ], the dosing cylinder is
filled automatically.
Parameter description
Dosing rate
Rate at which it is dosed.
Input range 0.01 to 166.00 mL/min
Selection maximum | dynamic
Default value maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 32, page 492).