29.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
296
■■■■■■■■
900 Touch Control
Measuring mode pH:
Input range –20.000 to 20.000
Selection off
Default value off
Measuring mode U, Ipol:
Input range –2000.0 to 2000.0 mV
Selection off
Default value off
Measuring mode Upol:
Input range
–200.0 to 200.0 µA
Selection off
Default value off
Start slope
Before starting the titration, it will be dosed until this slope (measured
value per volume) is reached. If the start slope is reached by the dosing of
a start volume, then the titration starts directly.
Measuring mode pH:
Input range 00.000 to 9,999 pH/mL
Selection off
Default value off
Measuring mode U, Ipol:
Input range 0 to 999 mV/mL
Selection off
Default value off
Measuring mode Upol:
Input range
0 to 99 µA/mL
Selection off
Default value off
Dosing rate
Rate to be dosed with until the start measured value or the start slope is
reached.
Input range 0.01 to 166.00 mL/min
Default value 5.00 mL/min
Selection maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 32, page 492).