29 Parameters
900 Touch Control
■■■■■■■■
305
Measured value (measuring mode U, Ipol):
Input range –2000.0 to 2000.0 mV
Default value –2000.0 mV
Measured value (measuring mode Upol):
Input range –200.00 to 200.00 µA
Default value –200.00 µA
Volume:
Input range 0.00000 to 9999.99 mL
Default value 0.00000 mL
Upper limit
Measured value or volume for the upper limit.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
instrument used.
Measured value (measuring mode pH):
Input range –20.000 to 20.000
Default value 20.000
Measured value (measuring mode U, Ipol):
Input range
–2000.0 to 2000.0 mV
Default value 2000.0 mV
Measured value (measuring mode Upol):
Input range
–200.00 to 200.00 µA
Default value 200.00 µA
Volume:
Input range
0.00000 to 9999.99 mL
Default value 9999.99 mL
EP criterion
The Equivalence point Recognition Criterion (ERC) found is compared with
this value. Equivalence points whose ERC is less than the value defined
here will be ignored.
Measuring mode pH:
Input range 0.10 to 9.99
Default value 0.50