29 Parameters
900 Touch Control
■■■■■■■■
341
Automatic start [Automatic start]
NOTICE
If you use the automatic start function, then we recommend that the
start drift be reduced and that a stabilizing time be defined.
Recommendation: start drift 10 µg/min / stabilizing time 30 s
Automatic start
on | off (Default value: off)
If this parameter is activated, the determination is automatically started
when the measured voltage suddenly changes. This setting will be ignored
as long as the working medium has not been conditioned.
Threshold value
If the change in voltage is higher than this value, the determination will
automatically be started.
Input range 0 to 999 mV
Default value 50 mV
Conditioning options [Cond. options]
Stabilizing time
Waiting time, during which the measured drift has to be smaller than the
start drift defined until Conditioning OK is displayed. The start drift is
defined in the dialog Edit command / Conditioning.
Input range 0 to 999999 s
Default value 0 s
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range 1 to 999999 s
Selection off
Default value off
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.