29 Parameters
900 Touch Control
■■■■■■■■
355
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 32, page 492).
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range 0 to 999999 s
Default value 0 s
Initial measured value
The [Initial meas. value] is measured before carrying out the start con-
ditions. This value is used for the titration direction determination
when the titration direction is set to auto.
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.
Input range 0.1 to 999.0 mV/min
Selection off
Default value off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range 0 to 999999 s
Default value 0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range 0 to 999999 s
Default value 1 s