29 Parameters
900 Touch Control
■■■■■■■■
385
Input range 0.00001 to 99999.9
Selection from list
Default value from list
from list
The standard solution selected under Standard addition ▶ Dosing
device is used.
Unit of the standard solution used.
Entry max. 24 characters
Selection mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L
Default value ppm
Number of additions
Number of standard additions. The more addition steps you carry out the
more accurate the evaluation becomes.
Input range 1 to 19
Default value 3
Stop volume
Maximum permissible volume for the volume steps. If the sum of the vol-
ume steps is higher than this volume, an error message is displayed when
starting the determination.
Input range 0.00 to 9999.90 mL
Default value 99.99 mL
Volume aux. solution
Volume of the buffer solution etc. which has to be added to the sample
before the measurement can be started.
Input range 0.00 to 9999.90 mL
Default value 0.00 mL
Stirrer control
The parameters for the stirrer control during the standard addition are
edited under [Stirrer control]. Ensure that a stirrer has been selected in
the Edit command / Stirrer dialog.
Stir solution during measurement
on | off (Default value: on)
If this parameter is activated, the stirrer will be switched on and off auto-
matically at the start or end of the measurement.