29 Parameters
900 Touch Control
■■■■■■■■
387
Selection Selection of configured titrants | not defined
Default value not defined
not defined
No check takes place.
Dosing rate
Three predefined sets of parameters can be selected for the dosing rate.
The maximum dosing rate depends on the cylinder volume (see chapter
32, page 492).
Selection slow | medium | fast
Default value fast
slow
The dosing rate is at 10 % of the maximum dosing rate.
medium
The dosing rate is at 50 % of the maximum dosing rate.
fast
The dosing rate is equivalent to the maximum dosing rate.
Delta U
This parameter can only be edited with Addition = auto.
The individual volume steps are calculated in the way that the voltage dif-
ference entered is reached.
Input range 1 to 999 mV
Default value 10 mV
Addition increments
The volume of the individual volume steps is defined under [Addition
increments]. This button is enabled only with Addition = auto dos or
manual.
Volume
Volume of a single volume step.
Input range 00.000 to 9999.900 mL
Default value 00.100 mL
Apply this value for all increments
on | off (Default value: off)
If you activate this parameter, the entered volume is applied for all volume
steps.