7.2 System-specific dialog options
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
900 Touch Control
Blocking commands
Proceed as follows to block unneeded commands:
1
Display the command list
Tap on the [Command list] button.
The list of all command groups is displayed.
2
Deactivate command groups
Deactivate those command groups which are not permitted to be
used.
3
Saving the settings
Tap on the fixed keys [ ] or [ ].
All deactivated commands appear grayed out in the method editor
and cannot be used for creating methods.
Blocking fixed keys
Proceed as follows to block unneeded fixed keys:
1
Display fixed keys which can be blocked
Tap on the [Fixed keys] button.
All fixed keys which can be blocked are displayed.
2
Deactivate fixed keys
Deactivate those fixed keys which are not permitted to be used.
3
Saving the settings
Tap on the fixed keys [ ] or [ ].
Deactivated fixed keys cannot be used.
Configuring the routine dialog
A suitable Standard configuration has already been saved for routine
operations:
Methods can only be loaded, but not modified or created.
Determinations cannot be recalculated.
You can readjust this standard configuration by disabling additional func-
tions or re-enabling disabled functions.