29 Parameters
900 Touch Control
■■■■■■■■
447
Rate at which it is dosed or filled.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum dosing and filling rates depend on the cylinder volume
(see chapter 32, page 492).
29.15.4.2 Control instrument
The control instrument the determination is carried out with is selected
under [Control device]. Control instruments are defined under Sys-
tem Device manager.
NOTICE
This button is only displayed when several control instruments have
been configured.
Control device
Selection of the control instrument from the list of devices. Only those
devices are displayed which are able to carry out the command. The con-
trol instruments do not have to be connected. This allows to transfer
methods more easily from one system to another.
Selection Selection of configured control instruments
29.15.4.3 Dosing device
The parameters for the dosing device are edited under [Dosing device].
Dosing device
Selection of the MSB connector to which the dosing device is connected.
Every MSB connector will always be displayed.
Selection 1 | 2 | 3 | 4
Default value 1
Titrant
Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, concentration, etc.)
is always used for the calculation. Titrants are defined under Sys-
tem Titrants. You can also enter a name which is not contained in the
titrant list. When a determination is started there is a check whether the
solution is contained in the list.