29.15 Dosing and Liquid Handling
■■■■■■■■■■■■■■■■■■■■■■
454
■■■■■■■■
900 Touch Control
29.15.5.2 Stop conditions
The conditions for canceling the dosing are defined under [Stop condi-
tions].
Stop volume
The dosing is stopped when the specified volume has been dosed since
the start of the dosing. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range 0.00000 to 99999.9 mL
Selection off
Default value off
Stop time
The dosing is stopped when the specified time has elapsed since the start
of the dosing.
Input range 0 to 999999 s
Selection off
Default value off
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 to 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see chapter
32, page 492).