32.10 Diagnosis
■■■■■■■■■■■■■■■■■■■■■■
522
■■■■■■■■
900 Touch Control
NOTICE
When updating to software version 5.900.0043, the memory will be
reformatted.
Before carrying out the software update, make a backup in order to
save your data and the system settings (see chapter 12.3, page 127).
Use a separate USB flash drive for the software update. It must be
removed from the instrument after the successful software update.
Program files
The files are instrument-specific. The file names are structured as follows:
■ Control software for 900 Touch Control:
5XXXyyyy.BIN
5XXXyyyy.MBIN
– XXX = instrument type (i.e. "900" for the 900 Touch Control)
– yyyy = program version
■ Firmware of the control instrument:
5XXXyyyy.BIN
– XXX = instrument type (e.g. 907 for the 907 Titrando)
– yyyy = program version
Language files
Language files may have the following content:
■ one or more additional dialog languages
■ the online help for one or more additional dialog languages
■ supplements for existing dialog languages/online help documents
They can be recognized by means of the two-digit language code in the
file name. The file name has the following structure:
■ 5XXXyZZZML.BIN
5XXXyZZZML.MBIN
– XXX = instrument type (i.e. "900" for the 900 Touch Control)
– y = version number of the language package
– ZZZ = program version