8.2 Editing titrant data
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
900 Touch Control
Unit of the titer.
Entry 10 characters maximum
Default value empty
Selection µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L
NOTICE
If you modify the titer or the concentration of the titrant in a loaded
determination at a later date and would then like to recalculate the
determination with the corrected value, then you must modify the value
in the determination data under View data / Titrant data (see "Cali-
bration and titrant data", page 218).
Date titer det.
Date and time of the last titer determination. For new titrants, the time
the preparation was made is specified until after the first time a titer deter-
mination has been carried out.
[Working life]
Defining the working life of the titrant (see chapter 8.3, page 45).
[Dosing unit]
This button is only displayed for Type = IDU or DU.
Defining the properties of the dosing unit used (see chapter 8.4, page
46).
[Exchange unit]
This button is only displayed for Type = IEU or EU.
Defining the properties of the exchange unit used (see chapter 8.5, page
52).
[Titer options]
Displaying the properties for titer determination (see chapter 8.7, page
58).