Appendix B ASAP 2460 Operator’s Manual
B-12 246-42800-01 - Dec 2013
2516- Warning: Keeping a backup copy of the calibration data is recommended by
Micromeritics. Would you like to do so now?
Cause: You have performed a calibration operation; a backup copy is recom-
mended.
Action: Perform a calibration
Save operation.
2517- Canceling this dialog will reset the calibration state to what it was when this
dialog was first opened. Are you sure you want to cancel?
Cause: You have not accepted the calibration you performed.
Action: If the calibration operation was successful, press
Accept.
2520- No data points available for reporting.
Cause: No analysis file.
Action: Choose another file.
2521- Unable to program controller.
Cause: A hardware malfunction has occurred.
Action: Contact your local Micromeritics service representative.
2522- Invalid controller application file.
Cause: The application’s control file has been corrupted or deleted.
Action: Reinstall the ASAP 2460 analysis program.
2523- Programming the controller failed.
2524- CRC check failed on programming controller.
2525- Unknown error programming controller.
Action: Calibration operations should only be done by or under the direction of
qualified service personnel.