ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-13
2526- Controller download was not successful.
2527- Controller CRC error on boot block.
2528- Controller DRAM error.
2529- Controller Com1: error.
2530- Controller Com2: error.
2531- Controller debug port error.
Cause: An internal processing and/or hardware error has occurred.
Action: Contact your service representative if you continue to receive this error
me
ssage.
2532- The instrument contains a different software version. Do you want to reset it?
Cause: The application has discovered a different version of software operat-
ing in the analyzer.
Action: If there are no analyzers other than the
ASAP 2460 connected to the
computer, select
Yes and allow the updated software to load.
2533- Analyzer initialization failed.
Cause: An internal processing and/or hardware error has occurred.
Action: Contact your service representative if you continue to receive this error
me
ssage.
2542- The file (filename) could not be written. Check that the directory permissions
are correct.
Cause: The person logged into the computer does not have write access to the
program and/or data directory.
Action: Have the system administrator of the computer set up proper privileges
for all users.
2548- System status 1 or 2.
Cause: A hardware error has occurred.
Action: Contact your service representative.