Level 2 Maintenance Training - M2000 Dispensing Systems 2-2-70 P/N 73-0018-00 (Revision B)
Exercise Sheet 2-9-2
Tactile Sensor Error
1. Gently lift the “hat” off of the tactile sensor.
2. Use isopropyl alcohol and a soft cloth to remove any residual fluid from around the
circumference of the “hat” and around the tactile sensor opening. Carefully remove
any residual fluid from the sensor tip as well.
3. Replace the “hat.
4. In the Main Window, click on Tools.
5. In the Tools Window, click on I/O Test.
6. Click on Dispenser.
7. Locate the I/O for tactile sensor.
8. Gently press down on the tactile sensor and make sure that the bit toggles ON
and OFF.
 If the bit does not toggle ON and OFF, the Needle Sensor may need to
be replaced. Contact Asymtek Technical Support.