Command reference
R&S
®
ZNL/ZNLE
1045User Manual 1178.5966.02 ─ 20
Example:
FREQ:MODE CW
Activate a time sweep.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
[SENSe<Ch>:]SEGMent<Seg>:CLEar
Deletes all sweep segments in the channel. The command is equivalent to
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. This suffix is ignored; the command
deletes all segments.
Usage: Event
[SENSe<Ch>:]SEGMent<Seg>:OVERlap
<Boolean>
Queries whether the analyzer supports overlapping sweep segments.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the command is instru-
ment-specific.
<Seg> Sweep segment number. This suffix is ignored; the command is
instrument-specific.
Parameters:
<Boolean> ON | OFF: No effect.
*RST: ON. If used as a query, the command returns the
information that overlapping sweep segments are
supported (ON).
[SENSe<Ch>:]SWEep:SPACing <StimulusFreq>
Defines the frequency vs. time characteristics of a frequency sweep ("Lin Frequency"
or "Log Frequency"). The command has no effect on segmented frequency or time
sweeps.
Note: Use [SENSe<Ch>:]SWEep:TYPE to select sweep types other than "Lin Fre-
quency" or "Log Frequency".
Suffix:
<Ch>
.
Channel number
Parameters:
<StimulusFreq> LINear | LOGarithmic
The stimulus frequency is swept in equidistant steps over the
frequency range. In a Cartesian diagram, the x-axis is a linear
frequency axis.
VNA command reference