VNA GUI reference
R&S
®
ZNL/ZNLE
509User Manual 1178.5966.02 ─ 20
File name / Comment (Optional)
Selects a filename to reference the characterization data set in the "Characterize Cal
Unit" and "Calibration Unit" dialogs and a comment, to be written into the characteriza-
tion file. A filename is required before you can "Finish" the characterization wizard and
store the data.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:CKIT
10.12.3 Power Cal Settings tab
R&S ZNL only
Gives access to all functions for power meter and power calibration data handling
(transmission coefficients). Power calibration is a fully menu-guided process.
Efficient power calibration procedure
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Power Cal Settings"
tab.
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to Chapter 9.5.6, "Scalar power calibration", on page 272.
Cal softtool