Command reference
R&S
®
ZNL/ZNLE
567User Manual 1178.5966.02 ─ 20
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage: Event
11.3 Common instrument commands
The following commands can be used independently of the active application(s).
● Commands for remote instrument operation.........................................................567
● Selecting the operating mode and application...................................................... 568
● Managing settings and results.............................................................................. 575
● Configuring the R&S ZNL/ZNLE........................................................................... 602
11.3.1 Commands for remote instrument operation
The following commands are required to shutdown or reboot the R&S ZNL/ZNLE from
a remote PC.
SYSTem:CLOGging....................................................................................................... 567
SYSTem:REBoot............................................................................................................567
SYSTem:SHUTdown...................................................................................................... 568
SYSTem:CLOGging <State>
This command turns logging of remote commands on and off.
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
Writes all remote commands that have been sent to a file.
The destination is C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\ScpiLogging\
ScpiLog.<no.>.
where <no.> is a sequential number
A new log file is started each time logging was stopped and is
restarted.
OFF | 0
*RST: 0
Manual operation: See "I/O Logging" on page 158
SYSTem:REBoot
This command reboots the instrument, including the operating system.
Common instrument commands