R&S ZNL Series - Calculate<Chn>:Limit:dcircle[:State]; Calculate<Chn>:Limit:delete:all; Calculate<Chn>:Limit:display[:State]

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
670User Manual 1178.5966.02 ─ 20
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> If set to ON, the line of the display circle is shown whenever the
related trace is displayed in complex format.
Manual operation: See "Show Border" on page 448
CALCulate<Chn>:LIMit:DCIRcle[:STATe] <Boolean>
Sets/queries the state of the display circle for the active trace of channel <Chn>.
The display circle is defined using CALCulate<Chn>:LIMit:DCIRcle:DATA.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> If set to ON, only trace points within the display circle are shown
at the GUI whenever the related trace is displayed in complex
format.
Manual operation: See "Limit to Circle On/Off" on page 448
CALCulate<Chn>:LIMit:DELete:ALL
Deletes all limit line segments.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 1.5 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 1.5 GHz, using default response values.
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment in the stimulus range
between 1.5 GHz and 2 GHz, assigning response values of +2
dB and +3 dB.
CALC:LIM:DEL:ALL
Delete both created limit line segments.
Usage: Event
Manual operation: See "Add / Insert / Delete / Delete All" on page 437
CALCulate<Chn>:LIMit:DISPlay[:STATe] <Boolean>
Displays or hides the entire limit line (including all segments) associated to the active
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace