EasyManua.ls Logo

R&S ZNL Series - Hcopy:abort; Hcopy:content

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
590User Manual 1178.5966.02 ─ 20
1 | ON
Logo is printed.
0 | OFF
Logo is not printed.
*RST: 1
Example:
DISP:LOGO OFF
Manual operation: See "Print Logo" on page 107
HCOPy:ABORt
This command aborts a running hardcopy output.
Example:
HCOP:ABOR
HCOPy:CONTent <Content>
This command determines the type of content included in the printout.
Parameters:
<Content> WINDows | HCOPy
WINDows
Includes only the selected windows in the printout. All currently
active windows for the current channel setup (or "MultiView") are
available for selection. How many windows are printed on a
each page of the printout is defined by HCOPy:PAGE:WINDow:
COUNt on page 596.
This option is not available when copying to the clipboard
(HCOP:DEST 'SYST:COMM:CLIP' or an image file (see
HCOPy:DEVice:LANGuage on page 593).
If the destination is currently set to an image file or the clipboard,
it is automatically changed to be a PDF file.
HCOPy
Selects all measurement results displayed on the screen for the
current channel setup (or "MultiView"): diagrams, traces, mark-
ers, marker lists, limit lines, etc., including the channel setup bar
and status bar, for printout on a single page. Displayed items
belonging to the software user interface (e.g. softkeys) are not
included. The size and position of the elements in the printout is
identical to the screen display.
*RST:
HCOPy
Common instrument commands

Table of Contents

Related product manuals