VNA GUI reference
R&S
®
ZNL/ZNLE
557User Manual 1178.5966.02 ─ 20
Note: The loaded file is stored in the active channel setup. Persisted channel setups
contain the full (de-)embedding data so that they can be transferred to other instru-
ments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<LogPt>
MMEMory:LOAD:VNETworks<Ch>:BALanced:EMBedding<LogPt>
10.14.7 Ground Loop tab
Allows you to specify a 1-port ground loop deembedding/embedding network.
Such a network is either defined
via its S-parameter stored in a one-port Touchstone file (*.s1p) or
by selecting a predefined lumped element model (Shunt L or Shunt C) and specify-
ing the available parameters (resistance/inductance or resistance/capacitance)
Use the complementary dock widget to activate or deactivate ground loop dembedding
and embedding (see Chapter 10.14.1.6, "Ground Loop panel", on page 539).
Background information
Refer to Chapter 9.6.2.7, "Ground loop de-/embedding", on page 290.