VNA GUI reference
R&S
®
ZNL/ZNLE
529User Manual 1178.5966.02 ─ 20
CONFigure:CHANnel<Ch>:NAME:ID?
CONFigure:CHANnel<Ch>:MEASure[:STATe]
INITiate<Ch>:CONTinuous
Add / Delete
The buttons below the channel table add and delete channels.
"Add" adds a new channel to the list. The new channel is named Ch<n>, where
<n> is the largest of all existing channel numbers plus one.
"Delete" deletes the channel selected in the table. This button is disabled if the
setup contains only one channel: In manual control, each setup must contain at
least one diagram area with one channel and one trace.
Remote command:
CONFigure:CHANnel<Ch>[:STATe]
10.13.2 Mode tab
Optimizes the measurement process.
10.13.2.1 Controls on the Mode tab
Figure 10-23: Channel > Mode tab (R&S
ZNL vs. R&S
ZNLE)
The controls on the "Mode" tab are organized in sections "Driving Mode" and "Image
Suppr." (R&S ZNL only).
Driving Mode
Determines the order of partial measurements and sweeps.
In "Auto" mode, the analyzer optimizes the display update: Fast sweeps are per-
formed in "Alternated" mode, slower sweeps in "Chopped" mode.
In "Alternated" mode, a partial measurement is performed at all sweep points (par-
tial sweep) before the hardware settings are changed. The next partial measure-
ment is carried out in an additional sweep.
This mode is usually faster than "Chopped" mode.
Channel softtool