7.5.3 Calibration signal display
Access: [Setup] > "Service" > "Calibration Signal"
Alternatively to the RF input signal from the front panel connector you can use the
instrument's calibration signal as the input signal, for example to perform service func-
tions on.
NONE .........................................................................................................................137
Calibration Frequency RF........................................................................................... 137
Frequency.....................................................................................................137
Calibration Frequency MW..........................................................................................138
NONE
Uses the current RF signal at the input, i.e. no calibration signal (default).
Remote command:
DIAGnostic:SERVice:INPut[:SELect] on page 606
Calibration Frequency RF
Uses the internal calibration signal as the RF input signal.
Remote command:
DIAGnostic:SERVice:INPut[:SELect] on page 606
DIAGnostic:SERVice:INPut:PULSed:CFRequency on page 606
Frequency ← Calibration Frequency RF
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration (max. 64 MHz).
Service functions