Command reference
R&S
®
ZNL/ZNLE
657User Manual 1178.5966.02 ─ 20
Example:
SWE:COUN 10
Define the number of sweeps (10) to be measured in single
sweep mode.
INIT:CONT OFF; :INIT; *OPC?
Activate single sweep mode and start a single sweep sequence
in channel no. 1. Wait until the single sweep sequence is com-
plete.
CALC:DATA:NSW? SDAT,3
Query the results of the 8
th
sweep.
See also "Retrieving the results of previous sweeps"
on page 1073.
Usage:
Query only
CALCulate<Ch>:DATA:SGRoup?
<Format>
Reads the current response values of all S-parameters associated to a group of logical
ports (S-parameter group). The S-parameter group must be created before using
CALCulate<Ch>:PARameter:DEFine:SGRoup.
Suffix:
<Ch>
.
Channel number of the previously defined S-parameter group.
Query parameters:
<Format> FDATa | SDATa | MDATa
Output format for the S-parameter data, see
CALCulate<Chn>:DATA on page 649.
Example: See CALCulate<Ch>:PARameter:DEFine:SGRoup
Usage: Query only
CALCulate<Chn>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See CALCulate<Chn>:DATA
Usage: Query only
CALCulate:DATA:TRACe <TraceName>, <Format>, <Data>...
The query gets the trace data of an arbitrary (not necessarily the active) trace, refer-
enced by its trace name <TraceName>.
The "set direction" allows to import formatted or unformatted trace data to an existing
trace.
Note
●
Unformatted trace data (SDATa) can only be imported to memory traces
VNA command reference